1Department of Internal Medicine, Kangwon National University School of Medicine, Chuncheon, Korea
2Department of Internal Medicine, School of Medicine, Kyungpook National University, Kyungpook National University Hospital, Daegu, Korea
3Department of Internal Medicine, Dongguk University, College of Medicine, Dongguk University Ilsan Hospital, Goyang, Korea
Copyright © 2018 Korean Society of Gastrointestinal Endoscopy
This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Author | Year | Actuation mode | Manipulation | n | Study design | Accuracy for gastric lesions | Safety (adverse events) | Note |
---|---|---|---|---|---|---|---|---|
Swain et al. [17] | 2010 | External magnetic | Hand-held magnet | 1 | Pilot study (case report) | N/A | none | First human trial |
Modified colon capsule (Given Imaging Ltd., Yoqneam, Israel) with external magnet paddle | ||||||||
Rey et al. [23] | 2010 | External magnetic | Robotic control | 53 | Pilot study (case series) | N/A | 0% | Magnetic driving system jointly developed by Olympus Medical Systems Corporation (Tokyo, Japan) and Siemens Healthcare (Erlangen, Germany) |
Keller et al. [16] | 2011 | External magnetic | Hand-held magnet | 10 | Pilot study (case series) | N/A | 0% | Modified colon capsule (Given Imaging Ltd.) with external magnet paddle |
Liao et al. [21] | 2012 | External magnetic | Robotic control | 34 | Pilot study (case series) | N/A | 0% | NaviCam (Ankon Technologies Co. Ltd., Wuhan, China) |
Rey et al. [4] | 2012 | External magnetic | Robotic control | 61 | Pilot study (single-center, prospective, self-controlled, blinded trial) | N/A | 3/61 (4.9%) | Magnetic driving system jointly developed by Olympus Medical Systems Corporation and Siemens Healthcare |
Jun et al. [6] | 2013 | Passive (no actuation) | Passive (no actuation) | 8 | Pilot study (case series) | 50%a) | 0% | PillCam ESO (Given Imaging Ltd.) with various position changes |
Denzer et al. [24] | 2015 | External magnetic | Robotic control | 189 | Two-center, prospective, self-controlled, blinded trial | 88.1%–90.5% | 0% | PillCam ESO (Given Imaging Ltd.) with innominate magnetic guidance system |
Zou et al. [27] | 2015 | External magnetic | Robotic control | 68 | Two-center, prospective, self-controlled, blinded trial | 91.2% | 3/68 (4.4%) | NaviCam (Ankon Technologies Co. Ltd.) |
Qian et al. [18] | 2016 | External magnetic | Robotic control | 60 | Case series | N/A | N/A | NaviCam (Ankon Technologies Co. Ltd.) |
Liao et al. [26] | 2016 | External magnetic | Robotic control | 350 | Multicenter, prospective, self-controlled, blinded trial | 93.4% | 5/350 (1.4%) | NaviCam (Ankon Technologies Co. Ltd.) |
Rahman et al. [14] | 2016 | External magnetic | Hand-held magnet | 26 | Pilot study (self-controlled comparative study) | 96%a) | N/A | MiroCam-Navi (Intromedic Ltd., Seoul, Korea) |
Author | Year | Actuation mode | Manipulation | n | Study design | Accuracy for gastric lesions | Safety (adverse events) | Note |
---|---|---|---|---|---|---|---|---|
Swain et al. [17] | 2010 | External magnetic | Hand-held magnet | 1 | Pilot study (case report) | N/A | none | First human trial |
Modified colon capsule (Given Imaging Ltd., Yoqneam, Israel) with external magnet paddle | ||||||||
Rey et al. [23] | 2010 | External magnetic | Robotic control | 53 | Pilot study (case series) | N/A | 0% | Magnetic driving system jointly developed by Olympus Medical Systems Corporation (Tokyo, Japan) and Siemens Healthcare (Erlangen, Germany) |
Keller et al. [16] | 2011 | External magnetic | Hand-held magnet | 10 | Pilot study (case series) | N/A | 0% | Modified colon capsule (Given Imaging Ltd.) with external magnet paddle |
Liao et al. [21] | 2012 | External magnetic | Robotic control | 34 | Pilot study (case series) | N/A | 0% | NaviCam (Ankon Technologies Co. Ltd., Wuhan, China) |
Rey et al. [4] | 2012 | External magnetic | Robotic control | 61 | Pilot study (single-center, prospective, self-controlled, blinded trial) | N/A | 3/61 (4.9%) | Magnetic driving system jointly developed by Olympus Medical Systems Corporation and Siemens Healthcare |
Jun et al. [6] | 2013 | Passive (no actuation) | Passive (no actuation) | 8 | Pilot study (case series) | 50% |
0% | PillCam ESO (Given Imaging Ltd.) with various position changes |
Denzer et al. [24] | 2015 | External magnetic | Robotic control | 189 | Two-center, prospective, self-controlled, blinded trial | 88.1%–90.5% | 0% | PillCam ESO (Given Imaging Ltd.) with innominate magnetic guidance system |
Zou et al. [27] | 2015 | External magnetic | Robotic control | 68 | Two-center, prospective, self-controlled, blinded trial | 91.2% | 3/68 (4.4%) | NaviCam (Ankon Technologies Co. Ltd.) |
Qian et al. [18] | 2016 | External magnetic | Robotic control | 60 | Case series | N/A | N/A | NaviCam (Ankon Technologies Co. Ltd.) |
Liao et al. [26] | 2016 | External magnetic | Robotic control | 350 | Multicenter, prospective, self-controlled, blinded trial | 93.4% | 5/350 (1.4%) | NaviCam (Ankon Technologies Co. Ltd.) |
Rahman et al. [14] | 2016 | External magnetic | Hand-held magnet | 26 | Pilot study (self-controlled comparative study) | 96% |
N/A | MiroCam-Navi (Intromedic Ltd., Seoul, Korea) |
N/A, not applicable. Not presented, but calculated from the results of manuscript.