Infectious transmission from Savary dilators is uncommon: retrospective review from a large tertiary academic center

Article information

Clin Endosc. 2022;55(5):691-694
Publication date (electronic) : 2021 June 29
doi : https://doi.org/10.5946/ce.2021.050
1Harvard Medical School, Boston, MA, USA
2Department of Medicine, Massachusetts General Hospital, Boston, MA, USA
3Division of Gastroenterology, Massachusetts General Hospital, Boston, MA, USA
Correspondence: Zoe N. Memel Massachusetts General Hospital, Gastroenterology Division, 55 Fruit Street, Boston, MA 02114, USA E-mail: zmemel@mgh.harvard.edu
Received 2021 January 18; Revised 2021 March 16; Accepted 2021 March 21.

In recent years, there have been numerous multi-drug resistant outbreaks attributed to endoscopes, with studies documenting residual bioburden on endoscopes, suggesting inadequate disinfection as the cause of these outbreaks.1,2 Wire-guided bougie dilation (WBD) is a common treatment strategy for esophageal strictures. Similar to gastrointestinal endoscopes, dilators have long, narrow lumens, and high-level disinfection (HLD) is used to reprocess these devices. However, despite the breadth of data on reprocessed endoscopes, there are no data on the prevalence of infectious transmission between reprocessed WBD dilators. Given the exposure of these instruments to the microbial bioburden in the oral cavity and esophagus, we sought to determine whether WBD dilation poses the same potential risk.

Based on the few studies performed on the reprocessing of flexible endoscopes, the increased level of pathogen transmission has been linked to inadequate cleaning, incorrect selection of high-level disinfecting agents, and/or improper drying and other breeches in reprocessing.3-5 Given the paucity of data evaluating the infectious transmission rate with reusable WBD dilators, the objectives of this study were (1) to determine the rate of infectious complications after use of WBD dilators during endoscopic procedures, and (2) to evaluate the etiology of readmission within 90 days following endoscopy with the use of WBD dilators.

Billing data from the Department of Gastroenterology, Massachusetts General Hospital were utilized to retrospectively identify patients who underwent endoscopy with WBD to treat esophageal strictures from 2000 to 2019. An infectious complication was defined as any infectious process that developed in a patient within 90 days after their endoscopic procedure. Patients who had esophageal stents placed during the same time period were excluded. Cases were examined for primary demographic data, immediate complication rates, and readmission within 90 days of the procedure with a review of all admission notes, emergency department visits, endoscopy reports, relevant progress notes, microbiology results, and discharge summaries. Data extracted included readmission etiology and possible confounding factors, including the recent use of immunosuppressive medications, antibiotics, chemotherapy, radiation, and any active infections at the time of the dilation. Reasons for readmission were grouped into infectious and non-infectious etiologies. The results were categorized into subgroups, and the percentage of each etiology for readmission was calculated. The bougie dilators evaluated in this study were manufactured by two manufacturers. Savary-Gilliard dilators were made by Cook Medical Device Company (Billerica, MA, USA), and American dilators were made by ConMed (Westborough, MA, USA). Our institutional protocol for WBD reprocessing utilizes ortho-phthalaldehyde for HLD; further information on the specific reprocessing protocol is detailed in Supplementary Material 1. This study was conducted in full accordance with all applicable Massachusetts General Hospital Policies and Procedures.

In total, 528 patients underwent WBD and met the inclusion criteria. Forty-four (8.3%) of these patients were readmitted within 90 days of the procedure (Table 1), of which eight had infections defined clinically or radiographically (18.2%). The most common infection was respiratory etiology (n=7, 15.9%), followed by gastrointestinal (n=1, 2.3%). Gastrointestinal (GI) infection was secondary to Clostridium difficile. The remaining 35 cases of readmission within the first 90 days (79.5%) were due to cardiac etiologies, non-infectious GI-related symptoms (nausea, vomiting, dysphagia), metabolic abnormalities (such as hyponatremia), and miscellaneous etiologies (including falls, alternate non-gastrointestinal procedures, etc.). The mean duration between procedure and readmission was 34.5 days.

Characteristics and outcomes of patients who had readmission to a hospital within 90 days after esophageal dilation using wire-guided bougie dilation

In reviewing the specific dilators used in each case, none of the same-sized dilators were used within a three-month period of another case with an infectious complication. Our Department of Gastroenterology has two complete sets of bougie dilators that range from 18 to 60 French within the endoscopy unit. We estimate that we average 1 dilation every 14 days and estimate that the same dilator is used every 30 days as a conservative approximation. In total, out of the patients readmitted within 90 days, four patients had undergone radiation within three months of the procedure, 11 patients had undergone both chemotherapy and radiation within three months of their procedure (25.0% of readmissions), and two patients had been taking an immunosuppressant agent, specifically mycophenolate mofetil or etanercept, prior to the procedure (4.5% of readmissions). Of the patients who underwent any form of chemotherapy, radiation, or immunosuppressive regimen, only two developed infections, both of which were respiratory in origin. Additionally, 7 patients (15.9%) had taken antibiotics within 3 months prior to the procedure. Of note, the patient diagnosed with C. difficile was not taking antibiotics prior to the procedure. Three patients had active infections at the time of their WBD dilations: one had a urinary tract infection, while two had chronic obstructive pulmonary disease exacerbations presumed to be due to respiratory tract infections. Patients with an infectious complication were readmitted on average 35.7 days after their procedure. To our knowledge, this is the first retrospective case series to assess the incidence of infectious complications and readmission after undergoing WBD. The findings of this study demonstrate that the 90-day readmission rate for infectious complications after WBD was low at approximately 8%. A small but finite risk of pulmonary and gastrointestinal infections was observed; however, based on analysis of the timeline of dilation among patients with consecutive infections, the likelihood of cross-contamination appears low. In previous endoscopy-associated transmission cases, transmission typically occurred within several weeks of contamination,1,2 and despite our 90-day window, the majority of cases occurred in the first 30 days, consistent with prior studies.The Spaulding classification system is utilized by the Centers for Disease Control and Prevention6 and the US Food and Drug Administration7 to categorize medical devices according to their risk of infection. A semi-critical device is one that contacts intact mucous membranes, such as endoscopes and WBDs. Devices within this class should receive high-level disinfection, defined as the destruction of all vegetative microorganisms, mycobacteria, small nonlipid viruses, medium lipid viruses, fungal spores, and some, but not all, bacterial spores.8,9 Although there remains a lack of consensus on the clinical value of routine microbiological monitoring of flexible endoscopes, there has been little discussion on the utility of monitoring the microbiological bioburden level on WBD. Current reprocessing guidance by the American Society of Gastrointestinal Endoscopy and the Association for the Advancement of Medical Instruments do not specifically address endoscopic dilators.9 However, the latter is anticipated to release a new informational, technical report, TIR99, specifically addressing the reprocessing of dilators, ultrasound, and manometry probes.10 The limitations of this study include its retrospective study design, lack of residual bioburden culture data, and inability to provide further information on dilator specifics (such as purchase date and number of uses). Given our inability to perform residual bioburden cultures, there is insufficient evidence to directly evaluate WBD transmission of infection.

In conclusion, the risk of infection-associated readmission within 90 days after WBD was low. Moving forward, well-designed studies measuring the incidence of pathogen-associated transmission during endoscopic procedures using WBD are needed to better understand the rate of infectious transmission and the efficacy of HLD in minimizing the residual bioburden on these instruments.

Supplementary Material

Supplementary Material 1. Bouige dilator and wire (non-disposable) reprocessing guide: Massachusetts General Hospital White 3 Endoscopy & Charles River gastrointestinal outpatient reprocessing rooms guidelines.

ce-2021-050-suppl.pdf

Supplementary material related to this article can be found online at https://doi.org/10.5946/ce.2021.050.

Notes

Conflicts of Interest

Yasmin Hernandez-Barco is Public analysis Inc in Behalf of Takeda, Nestle Health Sciences. The other authors have no potential conflicts of interest.

Funding

None.

Author Contributions

Conceptualization: all authors; Data curation: ZNM, DJF; Formal analysis: ZNM, DJF; Methodology: KK; Supervision: KK; Validation: KK; Writing-original draft: ZNM, DJF, KK; Writing-review & editing: all authors.

References

1. Wendorf KA, Kay M, Baliga C, et al. Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak. Infect Control Hosp Epidemiol 2015;36:634–642.
2. Kenters N, Huijskens EGW, Meier C, et al. Infectious diseases linked to cross-contamination of flexible endoscopes. Endosc Int Open 2015;3:E259–E265.
3. Rutala WA, Weber DJ, ; the Healthcare Infection Control Practices Advisory Committee (HICPAC). Guideline for disinfection and sterilization in healthcare facilities, 2008 [Internet]. Atlanta (GA): Centers for Disease Control and Prevention; 2008. [updated 2019 May 24; cited 2021 Apr 6]. Available from: https://www.cdc.gov/infectioncontrol/guidelines/disinfection/index.html.
4. Alfa MJ, Sepehri S, Olson N, et al. Establishing a clinically relevant bioburden benchmark: a quality indicator for adequate reprocessing and storage of flexible gastrointestinal endoscopes. Am J Infect Control 2012;40:233–236.
5. Bisset L, Cossart YE, Selby W, et al. A prospective study of the efficacy of routine decontamination for gastrointestinal endoscopes and the risk factors for failure. Am J Infect Control 2006;34:274–280.
6. Centers for Disease Control and Prevention (CDC). A rational approach to disinfection and sterilization [Internet]. Atlanta (GA): CDC; 2016. [updated 2016 Sep 18; cited 2021 Apr 6]. Available from: https://www.cdc.gov/infectioncontrol/guidelines/disinfection/rational-approach.html.
7. US Food and Drug Administration (FDA). Reprocessing medical devices in health care settings: validation methods and labeling [Internet]. Silver Spring (MD): FDA; 2015. [updated 2018 Mar 6; cited 2021 Apr 6]. Available from: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/reprocessing-medical-devices-health-care-settings-validation-methods-and-labeling/.
8. Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy 2007;39:175–181.
9. Reprocessing Guideline Task Force, Petersen BT, Cohen J, et al. Multisociety guideline on reprocessing flexible GI endoscopes: 2016 update. Gastrointest Endosc 2017;85:282–294.
10. The Association for the Advancement of Medical Instrumentation (AMMI). Standards Spotlight: AAMI TIR99 to provide much-needed guidance for reprocessing dilators and probes [Internet]. Arlington (VA): AMMI; 2018. [updated 2018 Oct 1; cited 2021 Apr 6]. Available from: https://www.aami.org/news/article/Standards-Spotlight-AAMI-TIR99-to-Provide-Much-Needed-Guidance-for-Reprocessing-Dilators-and-Probes-71335.

Article information Continued

Table 1.

Characteristics and outcomes of patients who had readmission to a hospital within 90 days after esophageal dilation using wire-guided bougie dilation

Characteristic Value
Total participants with readmission after EGD within 90 days 44 (8.3% of total participants, n total = 528)
Female 27 (61.4)
Male 17 (38.6)
Mean age (yr) 68.5
Average days to readmission (day) 34.5
Indication for esophagogastroduodenoscopy
 Benign esophageal stricture 33 (75.0)
 Malignant esophageal stricture 7 (15.9)
 Cervical web/Schatzki ring 4 (9.1)
Factors affecting patients’ risk for post-dilation complication
 Antibiotics within 3 months prior to procedure 7 (15.9)
 Active infection at time of procedure 3 (6.8)
 Immunosuppressant agent use at time of procedure 2 (4.5)
 Undergoing radiation at time of procedure 4 (9.1)
 Undergoing chemotherapy at time of procedure 0 (0)
 Undergoing chemotherapy & radiation at time of procedure 11 (25.0)
Readmission for infectious diagnoses
 Total infectious readmissions 8 (18.2)
 Respiratory infection 7 (15.9)
 Gastrointestinal infection 1 (2.3)
Readmissions for non-infectious diagnoses
 Total non-infectious readmission 35 (79.5)
 Cardiac diagnosis 7 (15.9)
 Gastrointestinal diagnosis 15 (34.1)
 Metabolic abnormality 5 (11.4)
 Miscellaneous diagnosis 12 (27.3)

Values are presented as number (%).