Aerosol extractor for COVID-2019 prevention during endoscopic procedure

Article information

Clin Endosc. 2022;55(6):815-818
Publication date (electronic) : 2022 August 29
doi : https://doi.org/10.5946/ce.2022.025
Department of Endoscopy, Fukushima Medical University Hospital, Fukushima, Japan
Correspondence: Takuto Hikichi Department of Endoscopy, Fukushima Medical University Hospital, 1 Hikarigaoka Fukushima, 960-1295, Japan E-mail: takuto@fmu.ac.jp
Received 2021 December 15; Revised 2022 January 9; Accepted 2022 January 14.

During the coronavirus disease 2019 (COVID-19) pandemic, peroral gastrointestinal (GI) endoscopy is considered a high-risk procedure for possible viral transmission. Several devices have been developed to prevent or remove aerosol droplets, such as masks and box-type droplet prevention devices.1-3 However, in the case of masks, aerosol droplets arising during scope insertion cannot be completely prevented. In addition, the box-type droplet prevention device can keep aerosol from the patient's mouth inside the box; however, upon removal of the box, the aerosols spread into the air, which is considered a problem.4 A transportable high-flow extractor (Free-100M; Forest-One Co., Tokyo, Japan), which has been used in the dental field, has been reported to be used for aerosol droplet control in the medical field.5 However, Free-100M has a large body and vacuum cup, as well as noise generated while vacuuming. Therefore, Free-100Next (Forest-One Co.) was developed as an improved version of Free-100M, as a novel transportable high-flow extractor. Free-100Next has a smaller body than Free-100M, and the vacuum cup is also smaller, with a diameter of 12 cm (Fig. 1A). The noise volume while vacuuming was also reduced by approximately 20% compared to the Free-100M. In addition, Free-100Next is equipped with a special filter called the ultra-low particulate air (ULPA) filter. This filter is guaranteed by the Japanese Industrial Standards to be able to trap more than 99.9995% of the particles with a diameter of 0.15 µm. Therefore, by using Free-100Next, more than 99.9% of the viruses are trapped by the ULPA filter, and the air used for suction escapes through the exhaust port of Free-100Next. Since ULPA filter is removable, it can be disposed after use with one patient and applying a new filter for the next patient of so that no COVID-19 remains in Free-100Next. Therefore, we believe that this device is suitable for GI endoscopy (Fig. 1B).

Fig. 1.

The appearance of Free-100Next (Forest-One Co., Tokyo, Japan). (A) An overall view of Free-100Next. The yellow arrow indicates the vacuuming part. (B) A scene of clinical practice wherein Free-100Next is used during upper gastrointestinal endoscopy. The vacuuming part (yellow arrows) is positioned near the mouth of the patient.

As proof of the efficacy of Free-100Next, an experiment using a mannequin showed whether Free-100Next can capture sprayed pseudo-saliva from the oral cavity of a mannequin (Fig. 2A). The vacuum power of Free-100Next can be set to nine levels from level 1 to level 9. As shown in the experimental video, the pseudo-saliva was successfully captured even at level 3 and was even more reliably captured at level 9 (Supplementary Video 1).

Fig. 2.

The vacuuming effect of Free-100Next (Forest-One Co., Tokyo, Japan). and its use with a box-type droplet prevention device. (A) The experiment was conducted in a dark room to verify the vacuuming effect of Free-100Next on pseudo-saliva using a mannequin. Pseudo-saliva was successfully captured using the Free-100Next. (B) Free-100Next was used with a box-type droplet prevention device. During the upper gastrointestinal endoscopy, Free-100Next was placed inside the AP Cover (ABIS Corp., Hyogo, Japan) from the patient’s head. Yellow arrows indicate the vacuuming parts in the Free-100Next.

However, fast-discharging aerosol droplets may not be completely captured by the Free-100Next. Therefore, it is best to use Free-100Next in conjunction with a box-type droplet prevention device (Fig. 2B, Supplementary Video 2). Most of the aerosols from the patient’s mouth were captured by Free-100Next, while the rest remained in the box. Free-100Next captured the aerosols remaining in the box.

To prove the vacuuming effect of Free-100Next, we conducted another experiment in which 0.025-µm-sized viruses (the size of SARS-CoV-2 is 0.1 µm) were circulated into a closed space of 25 m3, and Free-100Next was activated to capture viral particles (Fig. 3A). This experiment was performed at the Kitasato Research Center for Environmental Science (Kanagawa, Japan) because the experiment could not be performed in an endoscopy room. Compressed air was supplied from a compressor to the nebulizer containing the viral solution, and the viral solution was sprayed into the closed space at 0.2 mL/min for 10 minutes to float the virus. Subsequently, Free-100Next placed in the center of the closed space was set to level 9 for 60 minutes of continuous vacuuming of the virus. An impinger that collects the floating virus was fixed 50 cm below the Free-100Next. The number of airborne viruses was calculated from the liquid collected from the impinger. Similarly, as a control group, the number of floating viruses was calculated using the same method when Free-100Next was not used. The number of floating viruses in the space decreased by 60% after 15 minutes, 84% after 30 minutes, and 97% after 60 minutes, compared to the non-activation of Free-100Next. As a result, the number of floating viruses in the space dramatically decreased compared to that of the control (Fig. 3B). Therefore, in actual GI endoscopic practice, the risk of aerosol diffusion into the endoscopy room would be reduced not only during endoscopy, but also when the box-type droplet prevention device is removed.

Fig. 3.

An experiment to prove the ability of Free-100Next (Forest-One Co., Tokyo, Japan). (A) An experiment was conducted to determine the ability of Free-100Next to capture externally injected viruses in a closed space of 25 m3 using a control without Free-100Next to verify the reduction in the number of viruses in the space over time. The yellow arrow indicates the vacuumed part of Free-100Next. (B) The number of floating viruses in the space decreased by 60% after 15 minutes, 84% after 30 minutes, and 97% after 60 minutes, compared to the non-activation of Free-100Next. The blue line shows the changes in the number of floating viruses in the closed space, where the Free-100Next is activated. The red line indicates the same situation without Free-100Next activation.

The Free-100Next, a novel transportable high-flow extractor, has certain limitations. First, Free-100Next alone cannot completely suppress aerosol droplets during GI endoscopy. Therefore, it is necessary to use masks and box-type droplet prevention devices. Second, it is costly. However, in the current situation where the novel coronavirus infection is not yet under control, the number of patients with COVID-19 necessitating GI endoscopy may increase in the future. Therefore, considering the high cost of COVID-19 treatment, we believe that Free-100Next will contribute to infection control. In conclusion, Free-100Next is useful for COVID-19 prophylaxis during an endoscopic procedure.

Supplementary Material

Supplementary materials related to this article can be found online at https://doi.org/10.5946/ce.2022.025.

Supplementary Video 1. An experiment was conducted in a dark room (https://doi.org/10.5946/ce.2022.025.v001).

ce-2022-025-v001.mp4

Supplementary Video 2. Clinical use of Free-100Next with box-type droplet prevention device (https://doi.org/10.5946/ce.2022.025.v002).

ce-2022-025-v002.mp4

Notes

Conflicts of Interest

The authors have no potential conflicts of interest.

Funding

None.

Acknowledgments

We express our gratitude to Dr. Mika Takasumi, Dr. Yoshinori Okubo, and Dr. Takumi Yanagita at Fukushima Medical Univer­sity Hospital for their advice regarding this study. Additionally, we wish to express our deep appreciation to Mr. Shogo Matsuda and Mr. Kiyoshi Ikeda from Forest-One Co. (Tokyo) and Mr. Akira Okaue and Mr. Yoshihiko Takahashi from Kitasato Research Cen­ter for Environmental Science (Kanagawa) and for their support in this study.

Author Contributions

Conceptualization: TH; Data curation: TH; Formal analysis: TH; Investigation: TH, TK, RK, MH, JN; Methodology: TH; Project administration: TH; Resources: TH; Visualization: TH; Writing–original draft: TH; Writing–review & editing: all authors.

References

1. Hikichi T, Jun N, Hamada K, et al. A novel endoscopic mouthpiece for COVID-19 prevention. Clin Endosc 2022;55:160–162.
2. Kobara H, Nishiyama N, Masaki T. Shielding for patients using a single-use vinyl-box under continuous aerosol suction to minimize SARS-CoV-2 transmission during emergency endoscopy. Dig Endosc 2020;32:e114–e115.
3. Sagami R, Nishikiori H, Sato T, et al. Endoscopic shield: barrier enclosure during the endoscopy to prevent aerosol droplets during the COVID-19 pandemic. VideoGIE 2020;5:445–448.
4. Simpson JP, Wong DN, Verco L, et al. Measurement of airborne particle exposure during simulated tracheal intubation using various proposed aerosol containment devices during the COVID-19 pandemic. Anaesthesia 2020;75:1587–1595.
5. Saito T, Fujishiro A, Asai T. Aerosol extractor for airway management of COVID-19 patients. J Anesth 2021;35:323.

Article information Continued

Fig. 1.

The appearance of Free-100Next (Forest-One Co., Tokyo, Japan). (A) An overall view of Free-100Next. The yellow arrow indicates the vacuuming part. (B) A scene of clinical practice wherein Free-100Next is used during upper gastrointestinal endoscopy. The vacuuming part (yellow arrows) is positioned near the mouth of the patient.

Fig. 2.

The vacuuming effect of Free-100Next (Forest-One Co., Tokyo, Japan). and its use with a box-type droplet prevention device. (A) The experiment was conducted in a dark room to verify the vacuuming effect of Free-100Next on pseudo-saliva using a mannequin. Pseudo-saliva was successfully captured using the Free-100Next. (B) Free-100Next was used with a box-type droplet prevention device. During the upper gastrointestinal endoscopy, Free-100Next was placed inside the AP Cover (ABIS Corp., Hyogo, Japan) from the patient’s head. Yellow arrows indicate the vacuuming parts in the Free-100Next.

Fig. 3.

An experiment to prove the ability of Free-100Next (Forest-One Co., Tokyo, Japan). (A) An experiment was conducted to determine the ability of Free-100Next to capture externally injected viruses in a closed space of 25 m3 using a control without Free-100Next to verify the reduction in the number of viruses in the space over time. The yellow arrow indicates the vacuumed part of Free-100Next. (B) The number of floating viruses in the space decreased by 60% after 15 minutes, 84% after 30 minutes, and 97% after 60 minutes, compared to the non-activation of Free-100Next. The blue line shows the changes in the number of floating viruses in the closed space, where the Free-100Next is activated. The red line indicates the same situation without Free-100Next activation.