Review Endoscopic full-thickness resection of upper gastrointestinal tract: a review on closure techniques
Siew Fung Hauorcid, Shannon Melissa Chanorcid
Clinical Endoscopy 2026;59(2):171-181.
DOI: https://doi.org/10.5946/ce.2025.037
Published online: September 29, 2025

Department of Surgery, The Chinese University of Hong Kong, Hong Kong, Hong Kong

Correspondence: Shannon Melissa Chan Department of Surgery, The Chinese University of Hong Kong, Hong Kong E-mail: shannonchan@surgery.cuhk.edu.hk
• Received: February 4, 2025   • Revised: May 20, 2025   • Accepted: May 22, 2025

© 2026 Korean Society of Gastrointestinal Endoscopy

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (https://creativecommons.org/licenses/by-nc/4.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

  • 5,759 Views
  • 366 Download
  • 4 Web of Science
  • 3 Crossref
  • 2 Scopus
prev next
  • Endoscopic full-thickness resection has become more and more popular. One of the most important parts of this procedure is the closure of these full-thickness defects. Apart from conventional through-the-scope (TTS) clips, several different methods and devices have emerged as safe and efficacious in recent years. New clips include the anchor pronged TTS clips, dual-action tissue clips, and over-the-scope-clips. There are also new line or loop-assisted closure methods such as clip loop method, reopenable clip over line method, loop 9 method, and the internal-traction-assisted suspended closure method. New devices include the helical tacking system and endoscopic suturing device. This review article will discuss in details the usage of these different methods and available literature on comparison between the different closure methods.
Endoscopic full-thickness resection (EFTR) of upper gastrointestinal (GI) tract lesions has emerged as a minimally invasive treatment option with promising outcomes. GI stromal tumors (GISTs) are the most common indication for EFTR in the upper GI tract. Other subepithelial lesions that may be treated with EFTR include neuroendocrine tumors and leiomyomas, among others. EFTR was first described in 1998.1
This technique offers a valuable alternative for managing these lesions while minimizing the associated risks of more invasive surgical approaches. Subsequently, Zhou et al.2 translated the technique into clinical practice and reported 26 successful cases of EFTRs. Currently, this technique is further subclassified into exposed and non-exposed and tunneled and non-tunneled types. Nevertheless, the closure techniques following EFTR are crucial for ensuring optimal healing and preventing complications such as leakage and peritonitis. This article focuses on various closure techniques for endoscopic procedures, detailing their technical aspects, advantages, and disadvantages. It also includes a comparison of these methods according to available literature. By examining these different approaches, the article aims to provide insights into their effectiveness and application in clinical practice, helping to guide decision-making for optimal patient outcomes.
Through-the-scope (TTS) endoscopic clips were originally developed as hemostatic tools, but they are now often used for tissue approximation in defect or fistula closure. The first case series on EFTR reported in 2011 described the closure via TTS endoscopic clips only.2 Multiple generations of TTS endoscopic clips are commercially available. The first generation has not gained much popularity because of the need for clip assembly before use. In contrast, the second generation features pre-assembled endoclips, such as the HX-5L (Olympus), Resolution (Boston Scientific), and TriClip (Cook Endoscopy).3 However, studies have reported a high rate of clip dislodgement, up to two-thirds of them within the first week and 80% within two weeks. Additionally, the clips’ rotation lacked precision, and the HX-5L model did not allow for reopening for position adjustment.4 The third generation of TTS endoscopic clips such as QuickClipPro (Olympus), SureClip (Micro-Tech Endoscopy), Resolution 360 (Boston Scientific), and Instinct (Cook Endoscopy) addressed the issues found in the second generation. They featured the ability to close and reopen, enabling adjustments to the application site and allowing for predictable rotation for precise deployment.
It is important to note that different clips exhibit varying profiles, for example, different jaw spans and opening angles, and different clips pose unique characteristics. Wang et al.5 investigated and compared five commercially available TTS clips based on parameters such as rotatability, overshoot, open/close precision, and tensile/closure strength across different endoscopic positions. Their findings revealed that the Resolution 360 clip demonstrated the best rotatability, while the Dura Clip and SureClip showcased the most precise open/closing ability, and the Instinct Clip was noted for having the best closure strength. These highlighted the importance of selecting the appropriate clip based on specific procedural needs.
In the closure of full-thickness defects, the durability of the clip is crucial for maintaining sustained closure. A durable clip ensures that the closure remains intact over time, reducing the risk of complications and promoting optimal healing. Animal studies showed that the one-week clip retention rates were as follows: 100% for Instinct clips, 67% for Resolution clips, and 50% for QuickClip2 Long clips. By four weeks, retention rates were 60% for Instinct clips, 40% for Resolution clips, and 0% for QuickClip2 Long clips.6 Thus, Resolution clips and Instinct clips could be the preferred option in view of the durability of the clips. Ultimately, the selection of clips should be based on the characteristics of the clips, and the size, nature, and location of the defect.
Mantis clip (Boston Scientific) was one of the specially designed TTS endoscopic clips that were designed to manage larger defects (Fig. 1). TruGrip anchor prongs at the tip of the clip allow a hold-and-drag maneuver and, hence, easier tissue approximation compared to conventional clips (Fig. 2). Nishiyama reported successful closure of a defect measuring 25 mm following gastric endoscopic submucosal dissection (ESD), achieved in just 7 minutes.7 Several retrospective cohort studies have demonstrated high technical success rates for defect closures of up to 6 cm. For instance, Shiomi et al.8 reported a series of 52 cases of colorectal ESD, achieving a complete closure rate of 98.1%, with a median closure time of 8 minutes; the median defect size was 32 mm, with the largest measuring 62 mm. Similarly, Tada presented a series of 32 colorectal ESD cases with a complete closure rate of 96.9% and a median closure time of 7.9 minutes. The median sizes of the resection specimens and tumors were 32 mm, ranging from 17 mm to 100 mm.9 To date, there have been no cohort studies examining defect closure using clips with anchor prongs after EFTR in the upper GI tract. However, this clip may represent a valuable tool for defect closure following upper GI EFTR.
The dual-action tissue (DAT) clip (Micro-Tech Endoscopy) is a novel TTS endoscopic clip introduced in 2022 (Fig. 3A). This closure device features two arms operated by separate handles and a central fixed post (Fig. 3B). This design allows for the closure of one arm first to capture one edge of the defect, followed by the other arm before the clip is deployed.10 This mechanism ensures a secure grip on each edge and enhances tissue approximation, particularly in larger defects. A multicenter prospective cohort study conducted in 2024, which included 107 cases of colonic endoscopic resection, demonstrated a high complete closure rate of 96.3% when using DAT clips in conjunction with conventional TTS endoscopic clips. The median defect size in this study was 40 mm.11 These findings illustrate the high technical success rate of DAT clips for tissue approximation in cases with large defect sizes. It is important to note that the catheter for this clip has a width of 2.9 mm, making it incompatible with standard gastroscopes that have a 2.8 mm working channel. Consequently, it necessitates the use of an endoscope with a working channel of 3.2 mm or greater.
The over-the-scope clip (OTSC) (Ovesco Endoscopy) is an advanced endoscopic closure device designed to manage GI defects and hemorrhage. The system consists of a mounted OTSC, thread, thread retriever, and a hand wheel for clip release. The system size ranges from 8.5 mm to 14 mm. For the application, a twin gasping forceps is used to grasp each edge of the perforation into the cap, followed by the firing of the mounted clip. This device provides a strong force for tissue approximation with a long clip retention time. One study showed the median time of clip retention was 14.5 months.12 Guo reported a series of 23 patients who underwent EFTR of gastric subepithelial tumors (SETs) measuring 2 cm or less, followed by closure using the OTSC. The closure success rate was 100%, with an average closure time of 4.9 minutes, and no complications were reported. Postoperative follow-up endoscopy at six months revealed no detachment of the OTSCs.13 However, the dimensions of both the cap and the clip restricted the use of this technique to the closure of smaller defects. The American Society for Gastrointestinal Endoscopy guidelines of 2019 recommended the use of the OTSC for defects measuring 2 cm or less.14
The full-thickness resection device (FTRD; Ovesco Endoscopy) is an innovative integrated device designed for simultaneous closure and resection. It combines an OTSC with a snare preloaded at the tip of the cap. The target lesion is first marked and then pulled into the cap using a grasper or suction. The cap's depth is greater than that of a conventional OTSC, allowing it to accommodate more tissue. Following this, the clip is deployed, and the lesion is resected using the hot snare. This technique qualifies as non-exposed EFTR, which reduces the risk of peritoneal contamination and tumor seeding compared to that via exposed EFTR. However, similar to traditional OTSC application, it is indicated only for SETs measuring 2 cm or less.15 In a prospective study involving 68 patients with upper GI or colorectal SETs undergoing EFTR with the FTRD, the defect closure rate was 100%, with histologically complete resection (R0) achieved in 98.5% of cases. The mean procedure time was 39.6 minutes, while the mean defect closure time was 5 minutes. Eleven patients (16.2%) experienced procedure-related adverse events, including one case of postoperative bleeding, eight cases of fever that resolved with antibiotics, and two cases of local peritonitis that were managed with nasogastric tube decompression.16 Another prospective study involving 23 gastric SETs treated with EFTR using the FTRD reported a technical success rate of 89.7% and an R0 resection rate of 76%.17 Additionally, a retrospective study of 40 cases of gastric SETs (36 GISTs and four leiomyomas) utilizing the FTRD found only three instances of incomplete resection; these cases were subsequently identified as having fallen off together with the OTSC, resulting in an overall treatment success rate of 100%.18 Overall, the FTRD represents a promising tool for non-exposed EFTR of small gastric SETs.
The use of endoloops and metal clips to close a full-thickness gastric defect was first described by Shi et al.19 in 2013. In this technique, a detachable endoloop was placed at the defect, followed by sequential metal clip applications to anchor the endoloop to the edge of the defect. The endoloop was tightened to approximate the defect (Fig. 4). Two or more endoloops may be used in this procedure if the defect is large. In the initial retrospective study of 20 cases, the mean size of the lesions was 1.47 cm (0.4–3 cm). Although five patients developed fever post-operatively, they were all managed conservatively with antibiotics and proton-pump inhibitors, and none of them developed severe complications such as peritonitis. In another series of 51 patients with gastric SETs with a mean lesion size of 2.4 cm (1.3–3.5 cm) who underwent EFTRs and closure with the clip loop technique, there were no cases with severe complications such as delayed or massive bleeding, peritonitis, or abdominal abscesses. A contrast study on day three was arranged according to their study protocol, and there was no leakage or delayed gastric emptying detected.20 In another similar series of 68 patients with gastric GISTs with a median lesion size of 2.6 cm (2.0–3.5 cm) who underwent EFTR, the clip loop technique achieved a 100% complete closure rate. Two complications were reported: one case of Mallory-Weiss syndrome managed endoscopically and the other case of delayed bleeding managed conservatively. No other complications, including leakage, were observed, as confirmed by a contrast study performed on postoperative day 3.21 Overall, the clip loop technique represents a valuable method for closure of full-thickness defects.
Numerous innovative techniques utilize suture lines and endoloops to assist closure of large defects because of their cost-effectiveness and widespread availability.
The loop 9 technique employs a self-made suture loop with a slip knot, which was tightened to bring the opposite edges of a sizable defect together using two metal clips, facilitating the application of additional endoclips to complete the closure (Fig. 5). In a pilot study including 20 cases of GI defect closure (including mucosal and full-thickness defects) with a median defect size of 17.5 mm (10–55 mm), the complete closure rate was 100%, and the sustained closure rate was 90%.22
The line-assisted complete closure (LACC) method was originally used for the closure of large colorectal ESD defects but was later explored for its potential in gastric full-thickness defect closure. In this technique, the first metal clip is secured with a nylon suture and anchored to the apex of the defect. Subsequent metal clips are alternately fixed on opposite sides of the defect, along with the suture. By placing traction on the suture, the edges of the defect are brought together and apposed effectively (Fig. 6). A pilot study in a porcine model was conducted by Yamasaki et al.23 to evaluate the outcomes of this technique in gastric full-thickness defect closure. The initial complete closure rate was 100%. However, on postoperative day 4, all six cases experienced partial or complete dehiscence. None of the cases had peritonitis or fluid leakage as the defects were all covered by surrounding tissues such as the omentum or the liver, and hence, the clinical course was good in all cases. The authors suggested that improvement in the LACC technique was needed to achieve a more secure closure after EFTR. Nomura developed a reopenable clip over line method, a conceptually similar technique to successfully closing a full-thickness gastric defect, initially demonstrated in a porcine model.24 This technique was later applied in a human case report, successfully closing a 17 mm full-thickness gastric defect.25 Additionally, another case report described the closure of an inadvertent full-thickness perforation following esophageal ESD.26 The loop 10 technique is a modification of the LACC technique to avoid the use of endoscopic scissors to cut the suture upon completion of closure. Its use in closure of full-thickness defects was first described in a case report of closure of an inadvertent perforation during a re-do peroral endoscopic myotomy (POEM) procedure. The closure was confirmed to be complete and sustained on follow-up endoscopy and contrast study.27
Gong et al.28 described an internal traction-assisted suspended closure technique. Instead of using an endoloop, they utilized a self-made rubber band created from the finger sleeve of a sterile surgical glove. By anchoring one metal clip at one end of the defect with the rubber band and securing the other end to adjacent tissue, the defect was aligned and easily apposed, allowing for closure with conventional TTS clips (Fig. 7). In their pilot study, they reported 8 cases following gastric EFTR, achieving a 100% closure success rate. The median defect length was 3.25 cm (ranging from 2.5 to 9.0 cm), with a median width of 2.8 cm (ranging from 1.8 to 6.0 cm). The median closure time was 13 min, and no patients experienced leakage or other adverse events.
X-Tack Endoscopic HeliX Tacking System (Boston Scientific) is a novel tool approved for clinical use in GI defect closure since 2020 by the US Food and Drug Administration (FDA). Each device consisted of 4 HeliX Tacks attached to a 3-0 polypropylene suture, and an X-Tack handle for deployment (Fig. 8). To achieve tissue approximation, each HeliX tack is delivered through the working channel of the endoscope (channel diameter of ≥2.8 mm) and screwed into the tissue at the edge of the defect one by one. Eventually, by tightening the suture and application of a cinch, the defect can be approximated. More than one X-Tack System can be used, depending on the size of the defect to be closed (Fig. 9).
In a multicenter retrospective study published in 2022, 93 cases with the use of X-Tack in GI defect closure were investigated. Seventeen of the cases were full-thickness defects, which included EFTR, perforation, POEM, anastomotic dehiscence, and fistula. The median defect size was 41.6±19.4 mm. The technical success rate was achieved in 89.2%. Only two cases of adverse events occurred: one with abdominal pain post-operatively managed conservatively, and the other one with delayed hemorrhage requiring endoscopic hemostasis. The authors suggested that the X-Tack system was useful in the closure of large and irregularly shaped defects not amenable to conventional TTS endoscopic clips or suturing devices.29 Another multicenter retrospective study published in 2024 investigated the use of the X-Tack in 56 participants. In this study, the overall technical and clinical success rates were 92.9% and 75%, respectively. However, most of the cases in this study were fistula, leakage, and perforation repair with a relatively low success rate, which was expected. There were seven cases of POEM-site closure in this study, and the technical and clinical success rates were 100%. Cases of EFTR were not included in this study.30 The X-Tack System serves as a useful tool to close GI defects where the location is difficult to reach or difficult to close owing to its irregularity or size.
Currently, there are three endoscopic suturing devices that are FDA approved: the OverStitch Endoscopic Suturing System (Boston Scientific), the Incisionless Operating Platform (IOP, USGI Medical), and the Endomina plication system (Endo Tools Therapeutics). This passage will focus on the OverStitch device, drawing upon the author’s experience.
The OverStitch, OverStitch Sx, and OverStitch NXT endoscopic suturing systems (Boston Scientific) are novel devices to achieve secure GI defect closure. The OverStitch device was approved for clinical use by the FDA in 2010. It consists of an end cap that is mounted on the end of the scope, a needle driver handle, an anchor exchange catheter, and a tissue retraction helix device. The first-generation OverStitch is mounted on a double-channel therapeutic endoscope, while the second-generation OverStitch Sx was developed to be used on a single-channel endoscope. The OverStitch NXT is the latest version of this device, featuring enhanced capabilities that allow for extreme retroflexion of up to 210°. This advanced maneuverability enables access to challenging anatomical regions, such as the fundus, facilitating more effective suturing in difficult areas. It also consists of a tissue helix, which could be fully controlled by a solo operator (Fig. 10). When using the device, the needle driver is first closed to pierce through one edge of the defect. The needle is then caught by the anchor exchange catheter and is passed back to the needle driver to prepare for the next bite of the opposite edge of the defect. The above steps are repeated until a figure-of-eight configuration suturing is performed, followed by the application of a cinch. The defect can be closed step-by-step until complete closure is obtained by repeating the above procedure (Fig. 11). In the procedure, a tissue helix retraction device can be used to screw into the tissue and allow tissue grasping and hence a better bite by the needle driver (Fig. 12).
There are a handful of case series and reports on using endoscopic suturing to close full-thickness defects. Andalib presented a series of seven cases involving gastric GISTs that underwent full-thickness resection followed by endoscopic suturing. The mean defect size was 24 mm (10–55 mm), with no reported complications post-procedure.31 Additionally, an analysis of a prospective database from the European registry regarding the OverStitch device included 137 patients. Among these, endoscopic suturing was performed on 32 patients with mucosal defects, achieving a 100% clinical success rate. There were 22 patients with full-thickness perforation closed with the OverStitch suturing device, and the clinical success rate was 94.7%. These findings indicate that the OverStitch technology offers a reliable method for defect closure, demonstrating significant clinical benefits.32
While OverStitch-related complications are infrequent, adverse events such as bleeding, mucosal injury, perigastric fluid collection, leak, and perforation have been reported.33 The passage of the device carries a risk of mucosal injury or laceration to areas such as the pharynx and esophagus. Consequently, meticulous maneuvering of the device is important. In some cases, the use of an overtube may be required to facilitate the device's passage.33
Currently, there are no human studies that provide a direct head-to-head comparison of endoscopic closure methods for full-thickness defects. A pre-clinical study conducted by Walradt and colleagues evaluated various methods for closing full-thickness GI defects, focusing on closure strength, time, technical and clinical success, and the NASA-TLX stress score.34 The study compared five techniques: OverStitch, X-Tack, OTSC, Mantis clips, and conventional TTS clips (Resolution 360 clips). In total, 60 ex vivo closures were performed, with the OverStitch demonstrating the highest suture strength, while X-Tack exhibited the weakest. The necropsy findings revealed that endoscopic suturing, OTSC, and Mantis clips achieved 100% full-thickness tissue closure, whereas TTS clips succeeded in only 25% of cases and X-Tack in none. In terms of closure times, the TTS clip group (including Resolution 360 and Mantis clips) was the quickest to apply, while the OverStitch and X-Tack required the most time. The NASA-TLX scores indicated that the OverStitch and X-Tack were associated with higher levels of perceived effort, while Mantis clips received the lowest scores. In the live animal study, nine out of ten subjects (90.0%) survived for seven days. One animal, which was assigned to the conventional TTS clip closure group, did not survive owing to an unsuccessful procedure. Overall, the technical success rate was notable, with four out of 20 perforations deemed unsuccessful, two involving conventional TTS clips, one X-Tack, and one OverStitch.
Another pre-clinical study by Matsui et al.35 compared gastric defect closure using conventional TTS clips, the clip loop technique, and OTSC. The results demonstrated that OTSC exhibited significantly higher closure strength and leak pressure compared to the other methods. In contrast, there was no significant difference in closure strength between the TTS clips and the clip loop technique. The study emphasized the advantages of OTSC in achieving effective gastric defect closure. However, the study also acknowledged a limitation of the OTSC technique, noting that it is only limited to relatively small defects.
In general, small defects can be closed using conventional TTS clips only. The use of OTSC is also feasible, enabling one-step closure and resection, thereby shortening the procedure time. For oval-shaped or moderately sized defects, specially designed clips such as DAT clips or Mantis clips are required. The X-Tack system may also be utilized. Multiple devices may be necessary or used in combination to close a defect. In one study involving EFTR of lesions smaller than 50 mm, the conversion rate to laparoscopic closure was 10%, indicating the difficulty of closing larger defects with endoscopic clips only.36 Therefore, for larger defects, secure closure is best achieved through endoscopic suturing, depending on the availability of the suturing device and the expertise of the surgeon. Certain defect locations, such as the duodenum, pose challenges for endoscopic suturing owing to narrow spaces. An alternative method is clip-assisted or loop-assisted closure, such as the clip loop technique, although it may offer less robust closure force compared to endoscopic suturing.
Some authors recommend laparoscopic surgery for lesions located on the anterior or posterior walls of the stomach.36 This is because once the gastric wall is dissected and perforated, pneumoperitoneum occurs, impairing the stomach’s distensibility and causing poor visibility and difficulty in endoscopic manipulation. Endoscopic subserosal dissection (ESSD) has been suggested to overcome this problem. Wang et al. reported a case series of 16 cases where en bloc resection was achieved in 100% of cases with intact tumor capsules.37 Minor bleeding occurred in all cases, with endoscopic hemostasis achieved.
Conventional TTS clips, priced between US dollars (USD) $20 and $200 per unit, are generally considered the most cost-effective option because of their affordability and ease of use. DAT clips and Mantis clips, each priced at USD $230 and USD $255 per unit, respectively, are more expensive but may be necessary for larger defects. To manage costs, many clinicians incorporate conventional TTS clips alongside DAT or Mantis clips. OTSC, costing around USD $830 per unit, provides secure closure and are particularly effective for small defects. While they may be unpopular purely from a cost-effectiveness standpoint compared to TTS clips, their specialized design can make them the most economical choice when fewer clips are required. The OverStitch NXT, with the suture device itself priced at approximately USD $1,780, is the most expensive option. However, this may be essential for the secure closure of large defects that are not amenable to other methods. In addition to the suture device, one suture, helix, and cinch cost USD $105, USD $590, and USD $205, respectively. The X-Tack system costs around USD $1,385 per system, including 4 HeliX tacks. Estimating the number of systems required to close a defect is crucial for comparing costs with other techniques. The clip loop technique, employing endoscopic loops and conventional TTS clips, may be a cheaper alternative for larger defects. While it is an alternative for closing large defects, it may require many clips and potentially more than one loop, increasing overall costs.
It is important to note that these prices are estimated and can vary based on factors such as the supplier, the country, the brand of devices used, and how different devices are combined. Additionally, the choice of closure technique should consider not only the cost but also clinical outcomes, ease of use, and the specific characteristics of the defect. By carefully selecting the appropriate method, clinicians can achieve optimal results while managing costs effectively (Table 1).2,8,9,11,16-18,20,21,29-32
In summary, as endoscopic surgery advances, innovative tools and techniques will keep emerging. It is vital for endoscopists to understand the appropriate closure methods before performing EFTR procedures. This article has highlighted various endoscopic closure devices and techniques. Endoscopists should select the appropriate device according to the defect size, location, and personal experience.
Fig. 1.
Mantis clip with TruGrip anchor prongs at the tip of the clip.
ce-2025-037f1.jpg
Fig. 2.
(A–D) Application of the Mantis clip: one leg of the clip is first anchored onto one edge of the defect. The clip is then dragged to the opposite edge of the defect. The clip is then deployed successfully, grasping both edges of the defect.
ce-2025-037f2.jpg
Fig. 3.
(A) Dual-action tissue clip with two arms and a central post. (B) Two separate handles of the dual-action clip.
ce-2025-037f3.jpg
Fig. 4.
Clip loop technique (schematic diagram): a detachable endoloop was placed at the defect, followed by sequential metal clip applications to anchor the endoloop to the edges of the defect. The endoloop was tightened to approximate the defect.
ce-2025-037f4.jpg
Fig. 5.
Loop 9 technique (schematic diagram): a self-made suture loop with a slip knot was tightened to bring the opposite edges of a sizable defect together by two metal clips, facilitating the application of additional endoclips to complete the closure.
ce-2025-037f5.jpg
Fig. 6.
Line-assisted complete closure method (schematic diagram): the first metal clip is secured with a nylon suture and anchored to the apex of the defect. Subsequent metal clips are alternately fixed on opposite sides of the defect, along with the suture. By pulling the suture, the edges of the defect are brought together and apposed effectively.
ce-2025-037f6.jpg
Fig. 7.
Internal traction-assisted suspended closure technique (schematic diagram): by anchoring one metal clip at one end of the defect with the rubber band and securing the other end to adjacent tissue, the defect was aligned and easily apposed, allowing for closure with conventional through-the-scope clips.
ce-2025-037f7.jpg
Fig. 8.
(A, B) X-tack Endoscopic HeliX Tacking System consists of 4 HeliX tacks attached to a 3-0 polypropylene suture (Fig. 5A) and a handle (Fig. 5B).
ce-2025-037f8.jpg
Fig. 9.
(A–D) Application of X-Tack: each HeliX tack is screwed into the tissue at the edge of defect one by one (Fig. 6A–D). The suture is tightened to approximate the defect (Fig. 6E). The cinch is applied finally to keep the amount of tension of the suture. The rest of the defect can be closed with additional endoclips or another X-Tack system.
ce-2025-037f9.jpg
Fig. 10.
(A, B) OverStitch NXT endoscopic suturing device.
ce-2025-037f10.jpg
Fig. 11.
Overstitch application. (A) The needle driver is first closed during the introduction of the device. (B) The needle is mounted onto the needle driver. (C) A bite of tissue at the edge of the defect is made by closing the needle driver. (D) The needle is caught at the tip of the anchor exchange catheter. (E) Closing of the needle driver with the needle mounted onto the needle driver again.
ce-2025-037f11.jpg
Fig. 12.
(A, B) Tissue retraction HeliX device.
ce-2025-037f12.jpg
Table 1.
Comparison of pros and cons of different closure methods
Device Advantage Disadvantage Cost
Conventional TTS clips Easy to use Limited amount of tissue grasping USD $20–200 per clip
Complete closure rate up to 100% with no leakage in a case series2 Limited sustainability
Mantis clips Better sustainability and grasping force Still difficult for wide defect >3 cm USD $255 per clip
Complete closure rate (96.9%–98.1%)8,9
DAT clips Easy to apply Require at least 3.2 mm working channel diameter USD $230 per clip
Secure grasp of each edge of the defect
Complete closure rate 96.3% in a case series11
OTSC Robust closure Only for small defect 20 mm or less USD $830 per clip
High treatment success rate (89.7%–100%)16,17,18 Need for scope withdrawal for device mounting
Potentially lower R0 resection rate
Clip loop technique Easily available Uncertain robustness of closure force Variable depending on the no. and brand of clips
Complete closure rate up to 100% with no leakage in case series20,21
X-Tack system Easy to apply Uncertain robustness of closure force Suture device: USD $1,385 per unit
Technical success rate 89.2%–100%29,30 Cinch: USD $205 per unit
OverStitch system Secure closure The need to withdraw the scope for device mounting Suture device: USD $1,780 per unit
Full-thickness tissue grasping Bulky device Suture: USD $205 per unit
No size limit Limited retroflexion (overcome by the latest NXT version) HeliX Pro: USD $590 per unit
Clinical success rate 94.7%–100%31,32 More time consuming Cinch: USD $205 per unit
Steeper learning curve

Prices are quoted from Hong Kong in USD and may vary between countries.

TTS, through-the-scope; USD, United States dollar; DAT, dual-action tissue; OTSC, over-the-scope clip.

  • 1. Suzuki H, Okuwaki S, Ikeda K, et al. Endoscopic full-thickness resection (EFTR) and waterproof defect closure (ENDC) for improvement of curability and safety in endoscopic treatment of early gastrointestinal malignancies. Prog Dig Endosc 1998;52:49–53.
  • 2. Zhou PH, Yao LQ, Qin XY, et al. Endoscopic full-thickness resection without laparoscopic assistance for gastric submucosal tumors originated from the muscularis propria. Surg Endosc 2011;25:2926–2931.ArticlePubMedPDF
  • 3. Kantsevoy SV. Closure methods and devices. Tech Gastrointest Endosc 2019;21:33–37.Article
  • 4. Shin EJ, Ko CW, Magno P, et al. Comparative study of endoscopic clips: duration of attachment at the site of clip application. Gastrointest Endosc 2007;66:757–761.ArticlePubMed
  • 5. Wang TJ, Aihara H, Thompson AC, et al. Choosing the right through-the-scope clip: a rigorous comparison of rotatability, whip, open/close precision, and closure strength (with videos). Gastrointest Endosc 2019;89:77–86.ArticlePubMed
  • 6. Saxena P, Ji-Shin E, Haito-Chavez Y, et al. Which clip?: a prospective comparative study of retention rates of endoscopic clips on normal mucosa and ulcers in a porcine model. Saudi J Gastroenterol 2014;20:360–365.ArticlePubMedPMC
  • 7. Nishiyama N, Matsui T, Nakatani K, et al. Novel strategy of hold-and-drag clip closure with mantis-like claw for post-gastric endoscopic submucosal dissection defect of <30 mm. Endoscopy 2023;55(S 01):E1244–E1245.ArticlePubMedPMC
  • 8. Shiomi D, Tanabe M, Uragami N, et al. Clinical utility of a novel anchor pronged clip for mucosal defect closure after colorectal endoscopic submucosal dissection (with video). Endosc Int Open 2024;12:E1127–E1133.ArticlePubMedPMC
  • 9. Tada N, Tamai N, Ito M, et al. Novel reopenable clip with anchor prongs facilitates mucosal defect closure after colorectal endoscopic submucosal dissection: pilot feasibility study (with video). Dig Endosc 2024;36:1164–1170.ArticlePubMed
  • 10. Yang D, Kadkhodayan K, Arain MA, et al. Novel dual-action tissue through-the-scope clip for endoscopic closure. VideoGIE 2022;7:345–347.ArticlePubMedPMC
  • 11. Mohammed A, Gonzaga ER, Hasan MK, et al. Low delayed bleeding and high complete closure rate of mucosal defects with the novel through-the-scope dual-action tissue clip after endoscopic resection of large nonpedunculated colorectal lesions (with video). Gastrointest Endosc 2024;99:83–90.ArticlePubMed
  • 12. Wang XJ, Mahmoud T, Gonzalez-Izundegui D, et al. Natural history and retention time of over-the-scope clips in clinical practice. Tech Innov Gastrointest Endosc 2022;24:262–268.Article
  • 13. Guo J, Liu Z, Sun S, et al. Endoscopic full-thickness resection with defect closure using an over-the-scope clip for gastric subepithelial tumors originating from the muscularis propria. Surg Endosc 2015;29:3356–3362.ArticlePubMedPMCPDF
  • 14. ASGE Technology Committee, Aslanian HR, Sethi A, et al. ASGE guideline for endoscopic full-thickness resection and submucosal tunnel endoscopic resection. VideoGIE 2019;4:343–350.ArticlePubMedPMC
  • 15. Bauder M, Schmidt A, Caca K. Non-exposure, device-assisted endoscopic full-thickness resection. Gastrointest Endosc Clin N Am 2016;26:297–312.ArticlePubMed
  • 16. Guo JT, Zhang JJ, Wu YF, et al. Endoscopic full-thickness resection using an over-the-scope device: a prospective study. World J Gastroenterol 2021;27:725–736.ArticlePubMedPMC
  • 17. Meier B, Schmidt A, Glaser N, et al. Endoscopic full-thickness resection of gastric subepithelial tumors with the gFTRD-system: a prospective pilot study (RESET trial). Surg Endosc 2020;34:853–860.ArticlePubMedPDF
  • 18. Wang W, Liu CX, Niu Q, et al. OTSC assisted EFTR for the treatment of GIST: 40 cases analysis. Minim Invasive Ther Allied Technol 2022;31:238–245.ArticlePubMed
  • 19. Shi Q, Chen T, Zhong YS, et al. Complete closure of large gastric defects after endoscopic full-thickness resection, using endoloop and metallic clip interrupted suture. Endoscopy 2013;45:329–334.ArticlePubMed
  • 20. Ye LP, Yu Z, Mao XL, et al. Endoscopic full-thickness resection with defect closure using clips and an endoloop for gastric subepithelial tumors arising from the muscularis propria. Surg Endosc 2014;28:1978–1983.ArticlePubMedPDF
  • 21. Shi D, Li R, Chen W, et al. Application of novel endoloops to close the defects resulted from endoscopic full-thickness resection with single-channel gastroscope: a multicenter study. Surg Endosc 2017;31:837–842.ArticlePubMedPDF
  • 22. Inoue H, Tanabe M, Shimamura Y, et al. A novel endoscopic purse-string suture technique, “loop 9”, for gastrointestinal defect closure: a pilot study. Endoscopy 2022;54:158–162.ArticlePubMed
  • 23. Yamasaki Y, Ohmori M, Toyosawa J, et al. Modified line-assisted complete closure of the defect after gastric endoscopic full-thickness resection: a pilot study in porcine models. Endosc Int Open 2022;10:E609–E615.ArticlePubMedPMC
  • 24. Nomura T, Sugimoto S, Nakamura H, et al. Reopenable clip over line method for the closure of full-thickness defect after gastric endoscopic full-thickness resection. Endoscopy 2022;54(S 02):E808–E809.ArticlePubMedPMC
  • 25. Tani Y, Uedo N, Nomura T. Reopenable-clip over-the-line method for closure of gastric endoscopic full-thickness resection defect. Dig Endosc 2023;35:e85–e86.ArticlePubMed
  • 26. Asada Y, Shichijo S, Li JW, et al. Reopenable-clip over-the-line method for closure of large perforation during esophageal endoscopic submucosal dissection. Endoscopy 2023;55(S 01):E1103–E1104.ArticlePubMedPMC
  • 27. Inoue H, Shimamura Y, Fukuda M, et al. “Loop-10” line-assisted clip closure method: closure of perforation in re-do peroral endoscopic myotomy. VideoGIE 2023;8:186–188.ArticlePubMedPMC
  • 28. Gong Y, Wang J, Chen T, et al. New endoscopic closure technique, “internal traction-assisted suspended closure,” for GI defect closure: a pilot study (with video). Gastrointest Endosc 2024;99:186–192.ArticlePubMed
  • 29. Mahmoud T, Song LM, Stavropoulos SN, et al. Initial multicenter experience using a novel endoscopic tack and suture system for challenging GI defect closure and stent fixation (with video). Gastrointest Endosc 2022;95:373–382.ArticlePubMed
  • 30. Canakis A, Dawod SM, Dawod E, et al. Efficacy, feasibility, and safety of the X-tack endoscopic HeliX tacking system: a multicenter experience. J Clin Gastroenterol 2024;58:1052–1057.ArticlePubMed
  • 31. Andalib I, Yeoun D, Reddy R, et al. Endoscopic resection of gastric gastrointestinal stromal tumors originating from the muscularis propria layer in North America: methods and feasibility data. Surg Endosc 2018;32:1787–1792.ArticlePubMedPDF
  • 32. Maselli R, Palma R, Traina M, et al. Endoscopic suturing for GI applications: initial results from a prospective multicenter European registry. Gastrointest Endosc 2022;96:780–786.ArticlePubMed
  • 33. Yu JX, Schulman AR. Complications of the use of the OverStitch endoscopic suturing system. Gastrointest Endosc Clin N Am 2020;30:187–195.ArticlePubMed
  • 34. Walradt T, Szvarca D, Zarrella S, et al. New generation endoscopic closure devices for full-thickness defects: a comparative pre-clinical study (with videos). Surg Endosc 2025;39:649–656.ArticlePubMedPDF
  • 35. Matsui T, Kobara H, Nishiyama N, et al. Comparison of purse-string suture versus over-the-scope clip for gastric endoscopic full-thickness closure: traction and leak pressure testing in ex vivo porcine model. BMC Surg 2023;23:20.ArticlePubMedPMCPDF
  • 36. Abe N, Takeuchi H, Ohki A, et al. Comparison between endoscopic and laparoscopic removal of gastric submucosal tumor. Dig Endosc 2018;30 Suppl 1:7–16.ArticlePubMedPDF
  • 37. Wang A, Niu Q, Chen Y, et al. Efficacy and safety of endoscopic subserosal dissection treatment for gastrointetinal submucosal tumors in the upper gastrointestinal tract. BMC Surg 2024;24:301.ArticlePubMedPMCPDF

Figure & Data

REFERENCES

    Citations

    Citations to this article as recorded by  
    • Technische Fortschritte in der endoskopischen Resektion
      Sandra Nagl, Johannes Wießner, Katharina Beyer, Helmut Messmann
      Die Innere Medizin.2026; 67(1): 11.     CrossRef
    • Beyond EFTR Versus LECS: Toward Risk‐Adapted Management of Gastric Submucosal Tumors
      Yoshinori Morita
      Digestive Endoscopy.2026;[Epub]     CrossRef
    • A Novel Endoscopic Suture Anchor (ESA) Device for Full‐Thickness Defect Closure: A Comprehensive Ex Vivo and In Vivo Evaluation (With Video)
      Ya Qi Zhai, Jian Cong Feng, Zhen Yu Liu, Fang Yuan Xiang, Nan Shen, Hao Qi Zhai, En Qiang Linghu
      Journal of Digestive Diseases.2026; 27(7-8): 342.     CrossRef

    • Cite
      CITE
      export Copy Download
      Close
      Download Citation
      Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

      Format:
      • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
      • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
      Include:
      • Citation for the content below
      Endoscopic full-thickness resection of upper gastrointestinal tract: a review on closure techniques
      Clin Endosc. 2026;59(2):171-181.   Published online September 29, 2025
      Close
    • XML DownloadXML Download
    Figure
    • 0
    • 1
    • 2
    • 3
    • 4
    • 5
    • 6
    • 7
    • 8
    • 9
    • 10
    • 11
    Related articles
    Endoscopic full-thickness resection of upper gastrointestinal tract: a review on closure techniques
    Image Image Image Image Image Image Image Image Image Image Image Image
    Fig. 1. Mantis clip with TruGrip anchor prongs at the tip of the clip.
    Fig. 2. (A–D) Application of the Mantis clip: one leg of the clip is first anchored onto one edge of the defect. The clip is then dragged to the opposite edge of the defect. The clip is then deployed successfully, grasping both edges of the defect.
    Fig. 3. (A) Dual-action tissue clip with two arms and a central post. (B) Two separate handles of the dual-action clip.
    Fig. 4. Clip loop technique (schematic diagram): a detachable endoloop was placed at the defect, followed by sequential metal clip applications to anchor the endoloop to the edges of the defect. The endoloop was tightened to approximate the defect.
    Fig. 5. Loop 9 technique (schematic diagram): a self-made suture loop with a slip knot was tightened to bring the opposite edges of a sizable defect together by two metal clips, facilitating the application of additional endoclips to complete the closure.
    Fig. 6. Line-assisted complete closure method (schematic diagram): the first metal clip is secured with a nylon suture and anchored to the apex of the defect. Subsequent metal clips are alternately fixed on opposite sides of the defect, along with the suture. By pulling the suture, the edges of the defect are brought together and apposed effectively.
    Fig. 7. Internal traction-assisted suspended closure technique (schematic diagram): by anchoring one metal clip at one end of the defect with the rubber band and securing the other end to adjacent tissue, the defect was aligned and easily apposed, allowing for closure with conventional through-the-scope clips.
    Fig. 8. (A, B) X-tack Endoscopic HeliX Tacking System consists of 4 HeliX tacks attached to a 3-0 polypropylene suture (Fig. 5A) and a handle (Fig. 5B).
    Fig. 9. (A–D) Application of X-Tack: each HeliX tack is screwed into the tissue at the edge of defect one by one (Fig. 6A–D). The suture is tightened to approximate the defect (Fig. 6E). The cinch is applied finally to keep the amount of tension of the suture. The rest of the defect can be closed with additional endoclips or another X-Tack system.
    Fig. 10. (A, B) OverStitch NXT endoscopic suturing device.
    Fig. 11. Overstitch application. (A) The needle driver is first closed during the introduction of the device. (B) The needle is mounted onto the needle driver. (C) A bite of tissue at the edge of the defect is made by closing the needle driver. (D) The needle is caught at the tip of the anchor exchange catheter. (E) Closing of the needle driver with the needle mounted onto the needle driver again.
    Fig. 12. (A, B) Tissue retraction HeliX device.
    Endoscopic full-thickness resection of upper gastrointestinal tract: a review on closure techniques
    Device Advantage Disadvantage Cost
    Conventional TTS clips Easy to use Limited amount of tissue grasping USD $20–200 per clip
    Complete closure rate up to 100% with no leakage in a case series2 Limited sustainability
    Mantis clips Better sustainability and grasping force Still difficult for wide defect >3 cm USD $255 per clip
    Complete closure rate (96.9%–98.1%)8,9
    DAT clips Easy to apply Require at least 3.2 mm working channel diameter USD $230 per clip
    Secure grasp of each edge of the defect
    Complete closure rate 96.3% in a case series11
    OTSC Robust closure Only for small defect 20 mm or less USD $830 per clip
    High treatment success rate (89.7%–100%)16,17,18 Need for scope withdrawal for device mounting
    Potentially lower R0 resection rate
    Clip loop technique Easily available Uncertain robustness of closure force Variable depending on the no. and brand of clips
    Complete closure rate up to 100% with no leakage in case series20,21
    X-Tack system Easy to apply Uncertain robustness of closure force Suture device: USD $1,385 per unit
    Technical success rate 89.2%–100%29,30 Cinch: USD $205 per unit
    OverStitch system Secure closure The need to withdraw the scope for device mounting Suture device: USD $1,780 per unit
    Full-thickness tissue grasping Bulky device Suture: USD $205 per unit
    No size limit Limited retroflexion (overcome by the latest NXT version) HeliX Pro: USD $590 per unit
    Clinical success rate 94.7%–100%31,32 More time consuming Cinch: USD $205 per unit
    Steeper learning curve
    Table 1. Comparison of pros and cons of different closure methods

    Prices are quoted from Hong Kong in USD and may vary between countries.

    TTS, through-the-scope; USD, United States dollar; DAT, dual-action tissue; OTSC, over-the-scope clip.