A 63-year-old woman with diabetes mellitus and bronchial asthma presented with abdominal pain and a fever. Laboratory tests revealed elevated levels of inflammatory markers and hepatobiliary enzymes. Contrast-enhanced computed tomography (CT) revealed an impacted stone in the distal common bile duct (CBD) and multiple hepatic abscesses, consistent with acute cholangitis with secondary abscesses. Endoscopic retrograde cholangiopancreatography was performed. After endoscopic sphincterotomy, the stone was removed, and a 5-Fr double-pigtail plastic endoscopic retrograde biliary drainage (ERBD) stent was placed into the right intrahepatic bile duct (Fig. 1A). The infection improved, and the patient was discharged with the stent left in situ to prevent recurrent cholangitis before cholecystectomy.
Three months later, intraoperative laparoscopy revealed that the proximal end of the ERBD stent had penetrated the CBD wall and was protruding outside the bile duct (Fig. 1B). The tip was covered with serosa with no bile leakage. Postoperative CT showed no fistulae (Fig. 1C). Fluoroscopy revealed that the proximal end of the right hepatic duct, had migrated to the CBD (Fig. 2A). After stent removal, cholangiography revealed contrast pooling within the shallow wall depression, but no bile leakage (Fig. 2B). To prevent bile leakage, an endoscopic nasobiliary drainage tube was placed for one week. Follow-up imaging confirmed resolution (Fig. 2C, D).
Plastic ERBD stents are widely used,1 and complications can include stent occlusion, migration, and cholangitis.1,2 Migration occurs in 5% to 10% of cases.2,3 CBD wall perforation caused by a thin (5-Fr) plastic stent has not previously been reported. Possible mechanisms include chronic mechanical irritation of the bile duct wall by the stent tip, induced by respiratory and peristaltic motion. These forces may cause localized ischemia and erosion, resulting in delayed perforation.
This case highlights that even small-caliber plastic stents can lead to rare complications, warranting careful follow-up after placement.
Conflicts of Interest
The authors have no potential conflicts of interest.
Funding
None.
Author Contributions
Conceptualization: AK; Data curation: AK, MH; Formal analysis: AK, MH; Investigation: AK, YM, NI; Methodology: AK; Project administration: AK, MH, YM, NI, KS; Resources: AK; Supervision: MH, KS; Validation: AK, MH; Visualization: AK; Writing–original draft: AK; Writing–review & editing: all authors.
Fig. 1.(A) Endoscopic retrograde biliary drainage (ERBD) stent (5-Fr double-pigtail plastic stent) was placed into the right intrahepatic bile duct. (B) Intraoperative laparoscopic view revealing that proximal end of the ERBD stent (arrow) had migrated through the bile duct wall and partially protruding outside the duct. (C) Postoperative computed tomography showed the proximal end of the ERBD stent (arrow) protruding outside the bile duct wall. No fluid accumulation suggestive of bile leakage was identified.
Fig. 2.(A) Fluoroscopy confirmed that the proximal end of the stent had protruded outside the bile duct (arrow). (B) Cholangiography after endoscopic retrograde biliary drainage stent removal showing slight contrast leakage (arrow) at the site of penetration, without evidence of active bile leakage. (C) To prevent potential bile leakage, an endoscopic nasobiliary drainage tube (6-Fr pigtail) was placed. (D) Follow-up cholangiography (one week later) showed no further contrast leakage.
REFERENCES
- 1. Dumonceau JM, Tringali A, Papanikolaou IS, et al. Endoscopic biliary stenting: indications, choice of stents, and results: European Society of Gastrointestinal Endoscopy (ESGE) clinical guideline-updated October 2017. Endoscopy 2018;50:910–930.ArticlePubMed
- 2. Johanson JF, Schmalz MJ, Geenen JE. Incidence and risk factors for biliary and pancreatic stent migration. Gastrointest Endosc 1992;38:341–346.ArticlePubMed
- 3. Katsinelos P, Kountouras J, Paroutoglou G, et al. Migration of plastic biliary stents and endoscopic retrieval: an experience of three referral centers. Surg Laparosc Endosc Percutan Tech 2009;19:217–221.ArticlePubMed
Citations
Citations to this article as recorded by